Ask An Expert

We will Follow up with you within 24 hours.

Touch Screen Thermal Cycler (Basic) LMTC-A106

Catalog

Specifications

Block Capacity 96 wells x 0.2 mL
Block Temperature Range 4 °C - 99.9 °C
Temperature Accuracy ≤ ± 0.1 °C (55 °C) ; ≤ ± 0.2 °C (≥ 90 °C)
Temperature Uniformity ± 0.2 °C
Ramping Rate 0.1 °C - 4 °C
Heating Rate 1.8 °C/s
Cooling Rate 1.8 °C/s
Hot-lid Temperature Range 30 °C - 110 °C
Max No. of Program More than 250
Max No. of Cycles 99
Increment / Decrement Time 0 - 9 min 59 sec (Long PCR)
Increment / Decrement Temperature 0.1 °C - 9.9 °C (Touchdown PCR)
Display 6.5” LCD Color Touch Screen
Data Output Port LAN
Power Consumption 600 W
Power Supply AC 110 V / 220 V ± 10 %; 50 / 60 Hz
Gross Weight 12.5 Kg

Pricing

Product Price Unit Action
Touch Screen Thermal Cycler (Basic) LMTC-A106 $6,684.00 Pack of 1

Features

Touch Screen for smooth operation Real time observation on 6.5” LCD color display Peltier based heating and cooling control SOAK function helps maintain annealing temperature Stepless adjustable Hot-lid to avoid evaporation Auto shut off when temperature is ˂ 30 °C or after cycle Auto pause / start during technical errors Block / Tube temperature control mode

Similar Products

Touch Screen Thermal Cycler (Gradient) LMTC-B103

  • Block Capacity: 96 x 0.2 mL , 12 x 8 PCR Strip
  • Block Temperature Range: 4 °C - 99.9 °C
  • Temperature Accuracy: ≤ ± 0.1 °C (50 °C) ; ≤ ± 0.2 °C (≥ 90 °C)
  • Temperature Uniformity: ± 0.3 °C
  • Ramping Rate: 0.1 °C - 4 °C

Touch Screen Thermal Cycler (Gradient) LMTC-B100

  • Block Capacity: 96 x 0.2 mL , 12 x 8 PCR Strip
  • Block Temperature Range: 4 °C - 99.9 °C
  • Temperature Accuracy: ≤ ± 0.1 °C (55 °C) ; ≤ ± 0.2 °C (≥ 90 °C)
  • Temperature Uniformity: ± 0.2 °C
  • Ramping Rate: 0.1 °C - 4 °C

Touch Screen Thermal Cycler (Basic) LMTC-A108

  • Block Temperature Range: 0 °C - 100 °C
  • Temperature Accuracy: ≤ ± 0.2 °C (35 °C - 100 °C)
  • Temperature Uniformity: ± 0.2 °C (95 °C)
  • Ramping Rate: Adjustable
  • Heating Rate: 3.5 °C/s
Touch Screen Thermal Cycler (Basic) LMTC-A102
  • Block Capacity: 32 wells x 0.2 mL
  • Block Temperature Range: 4 °C - 100 °C
  • Temperature Accuracy: ± 0.25 °C
  • Temperature Uniformity: ± 0.25 °C
  • Ramping Rate: Adjustable
Touch Screen Thermal Cycler (Basic) LMTC-A101
  • Block Capacity: 16 wells x 0.2 mL
  • Block Temperature Range: 4 °C - 100 °C
  • Temperature Accuracy: ± 0.25 °C
  • Temperature Uniformity: ± 0.25 °C
  • Ramping Rate: Adjustable

Touch Screen Thermal Cycler (Gradient) LMTC-B101

  • Block Capacity: 96 x 0.2 mL , 12 x 8 PCR Strip
  • Block Temperature Range: 4 °C - 99.9 °C
  • Temperature Accuracy: ≤ ± 0.1 °C (50 °C) ; ≤ ± 0.2 °C (≥ 90 °C)
  • Temperature Uniformity: ± 0.2 °C
  • Ramping Rate: 0.1 °C - 4 °C

Touch Screen Thermal Cycler (Basic) LMTC-A100

  • Block Capacity: 16 wells x 0.2 mL
  • Block Temperature Range: 0.1 °C - 99.9 °C
  • Temperature Accuracy: ± 0.25 °C
  • Temperature Uniformity: ± 0.25 °C
  • Ramping Rate: Adjustable

Touch Screen Thermal Cycler (Gradient) LMTC-B104

  • Block Capacity: 384 wells plate
  • Block Temperature Range: 4 °C - 99.9 °C
  • Temperature Accuracy: ≤ ± 0.1 °C (55 °C) ; ≤ ± 0.2 °C (≥ 90 °C)
  • Temperature Uniformity: ± 0.2 °C
  • Ramping Rate: 0.1 °C - 4 °C
Touch Screen Thermal Cycler (Basic) LMTC-A103
  • Block Capacity: 32 wells x 0.2 mL
  • Block Temperature Range: 0.1 °C - 99.9 °C
  • Temperature Accuracy: ± 0.25 °C
  • Temperature Uniformity: ± 0.25 °C
  • Ramping Rate: Adjustable
Touch Screen Thermal Cycler (Gradient) LMTC-B105
  • Block Temperature Range: 0 °C - 100 °C
  • Temperature Accuracy: ≤ ± 0.2 °C (35 °C - 100 °C)
  • Temperature Uniformity: ± 0.2 °C (95 °C)
  • Ramping Rate: Adjustable
  • Heating Rate: 3.5 °C/s

Touch Screen Thermal Cycler (Gradient) LMTC-B107

  • Block Temperature Range: 0 °C - 100 °C
  • Temperature Accuracy: ≤ ± 0.2 °C (35 °C - 100 °C)
  • Temperature Uniformity: ± 0.2 °C (95 °C)
  • Ramping Rate: Adjustable
  • Heating Rate: 3.5 °C/s

Touch Screen Thermal Cycler (Basic) LMTC-A107

  • Block Temperature Range: 0 °C - 100 °C
  • Temperature Accuracy: ≤ ± 0.2 °C (35 °C - 100 °C)
  • Temperature Uniformity: ± 0.2 °C (95 °C)
  • Ramping Rate: Adjustable
  • Heating Rate: 3.5 °C/s

Touch Screen Thermal Cycler (Basic) LMTC-A109

  • Block Temperature Range: 0 °C - 100 °C
  • Temperature Accuracy: ≤ ± 0.2 °C (35 °C - 100 °C)
  • Temperature Uniformity: ± 0.2 °C (95 °C)
  • Ramping Rate: Adjustable
  • Heating Rate: 5 °C/s
Touch Screen Thermal Cycler (Gradient) LMTC-B200
  • Block Capacity: 6 x 16 wells x 0.2 mL , 96 wells x 0.2 mL , 12 x 8 PCR strips
  • Block Temperature Range: 4 °C - 100 °C
  • Temperature Accuracy: 4 °C - 100 °C
  • Temperature Uniformity: ≤ ± 0.15 °C (≥ 90 °C)
  • Ramping Rate: 0.1 °C - 4 °C
Touch Screen Thermal Cycler (Basic) LMTC-A200
  • Capacity: 96
  • Reaction Volume: 10 to 200ul
  • Tube Type: 96×0.2ml PCR plate, 8×0.2ml PCR tube
  • Block Temperature Range: 4°C to 105°C
  • Heat Lid Temperature Rangs: 30°C to 110°C