Overview
TOC Analyzer LMTOC-B104 offers automatic sample dilution and 0 to 1000 mg/L measurement range. It features 680°C catalytic oxidation with platinum catalyst for reliable measurement. Equipped with three-stage dehydration system, it enhances moisture removal and improves final gas drying. Incorporates standard curve management that simplifies calibration and enhances convenience. Our Analyzer is ideal for TOC measurement in seawater, pharma, industrial, and environmental samples.
Features :
Multi-functional software
PID temperature control
Consumables alert
Modular design
High sensitivity
Simplified operation
Specifications :
| Measurement Range | 0 to 1000 mg/L, can extend to 0 to 100,000mg/L |
| Measurement Method | High Temperature Catalytic Combustion |
| Detector | NDIR |
| Detection Limit | IC: 20μg/L |
| Analysis Parameter | TC, TIC, TOC, NPOC |
| Control Mode | PC software Control |
| Gas Requirement | Oxygen, purity greater than equal to 99.995% |
| Sample Type | Solid Sample |
| Dilution Method | Automatic |
| Measuring Time | IC: 3min |
| Maximum Permissible Error | IC: ± 4% |
| Repeatability | Less than or equal to 3% |
| Injection Volume | IC: 100 to 2000μL |
| Power Consumption | 700W |
| Power Supply | AC110/220V, 50/60Hz |
| Dimension | 450×500×480 mm |
| Packing Dimension | 780×710×610 mm |
| Net Weight | 46 Kg |
| Gross Weight | 66 Kg |
Optional Accessories :
Model No. | Accessory Name |
LMTOC-B104-A1 | Autosampler (Liquid) |
LMTOC-B104-A2 | Autosampler (Solid) |
Applications :
TOC Analyzer LMTOC-B104 is ideal for measuring TOC in seawater, pharmaceutical-grade water, industrial wastewater, and environmental samples to ensure compliance, process control, and high-purity water quality assurance.