Close
Quick Login
Close
Enquiry Form
Close
Ask An Expert
Overview

Ion Chromatography LMICS-A100 is designed with a bipolar conductivity detector to improve instrument detection performance and stability. Boasts a modular design coupled with advanced circuit technology. It provides a flow range spanning from 0.001 to 9.999 ml/min and has a detection range spanning from 0 to 35,000 µS/cm. It is exclusively crafted for the separation, identification, and quantification of ions within a sample.

Specifications :
Ion Chromatographic Pump
Type High pressure and low pulse double plunger tandem advection pump
Maximum pressure 35 MPa (PEEK)
Pressure display accuracy (0.2 to 0.5) ml/ min ≤ 3 %
(0.5 to 1.0) ml/ min ≤ 2 %
> 1.0 ml/ min 2 %
Flow stability 0.001 to 9.999 ml/ min
Flow range 0.001 to 9.999 ml/ min
Pressure pulse ≤ 0.5%
Conductivity detector
Type Temperature Control and Bipolar Conductivity Detector
Detection mode Bipolar conductivity detection
Cell volume ≤ 0.8 µL
Detection range 0 to 35000 µS/ cm
Detection resolution ≤ 0.0020 nS/ cm
Output voltage -6000 to +6000 mV (adjustable)
Baseline noise ≤ 0.05 % F.S.
Baseline drift ≤ 3 % FS/ 30 min
Operating temperature range Room temperature +5 to 60°C (41 to 140°F)
Controlling temperature accuracy ± 0.01°C
Maximum pressure 10 MPa
Linear range ≥ 103
Instruments linearity ≥ 0.999
Quantitative repeatability ≤ 1.0 %
Qualitative repeatability ≤ 1.0 %
Minimum detectable concentration Cl ≤ 0.0005 ug/ ml
BrO3 ≤ 0.001 ug/ ml
Li+ ≤ 0.001 ug/ ml
Flow system
Plastic flow path Made of PEEK materials
Six-way Valve PEEK material, 5000 psi pressure, independent automatic collection, and flow function
Degassing Device
Inner volume 300 µl
Degassing volume 7.5 ml
Maximum flow rate 10 ml/ min
Degassing efficiency Degassing 90 % at 1ml/ min
Organic tolerance 1
Built-in Eluent Generator
Eluent types KOH/ NaOH
Eluent concentration range 0.1 to 100 mM
Concentration increment 0.1 mM
Flow rate change 0.1 to 3.0 ml/ min
Maximum pressure 20 MPa
Minimum pressure 5 MPa
Power supply 150 W
Dimension 350 × 470 × 650 mm
Net weight 34 kg
Gross weight 40 kg
Features :
  • -  Featuring a gradient elution system, efficiently detects small molecules in complex samples
  • -  Incorporate temperature control for optimal analysis conditions
  • -  Includes automated components and software to streamline analysis
  • -  Built-in low-pressure degassing technology to remove bubble interference
  • -  Automated self-regenerating electrolytic microfilm suppressor
  • -  Enhanced with standard dilution function that saves time, labor, and costs
  • -  RS232 and USB interfaces
Standard Accessories :

Ion Chromatography host:1


High-pressure advection pump:1
Conductivity detector:1
Anion column:1
Anion guard column:1
Suppressor:1
Chromatography workstation:1
Column thermostat system:1
Spare parts kits :1
Vacuum degassing pump:1
Sand core filtration devices:1
Pretreatment column:20
0.22 µm syringe filter:200
Eluent generator:1

Optional Accessories :

Autosampler (20 samples)


Autosampler (54 samples)
Autosampler (108 samples)
Anion column
Guard column
Anion suppressor
Spare parts kits

Applications :

Ion Chromatography finds application in various scientific and industrial fields such as Environmental Monitoring, Pharmaceutical Analysis, Research Laboratories, and more.