Split Beam UV-Vis Spectrophotometer LMUS-D100

Split Beam UV-Vis Spectrophotometer LMUS-D100 operates across 190 to 1100 nm wavelength range for versatile spectrum measurements. Features split-beam optical system integrated with 1200 lines/mm grating for stable baseline correction. Engineered for high-performance absorbance and transmittance evaluation in scientific and industrial settings. Designed with 2nm spectral bandwidth suitable for routine and research-grade testing. Our Spectrophotometer incorporates deuterium and tungsten lamp. for extended UV and visible light coverage.

Catalog Send Mail Whatsapp

Specifications

Wavelength 190 nm to 1100 nm
Spectral Bandwidth 2 nm
Optical System Split Beam, Grating 1200 lines/mm
Wavelength Accuracy ±0.1nm at 656.1nm, ±0.3nm at other wavelengths
Wavelength Repeatability ≤0.1 nm
Photometric Range 0 to 400%T, –4 to +4A
Photometric Accuracy ±0.2%T(0 to 100%T), ±0.0005A(0 to 0.5A), ±0.0008A(0.5 to 1A)
Photometric Repeatability ≤0.15%T (0 to 100%T), 0.0004A(0 to 0.5A), 0.0006A(0.5 to 1A)
Resolution 0.00001Abs
Stray Light ≤0.03%T at 220nm, 360nm
Stability ±0.0001A
Baseline Flatness ±0.0004A
Noise 0.0005A at 500nm
Detector Imported Silicon Photodiode
Light Source Deuterium and Tungsten lamp
Work Mode T, A, C, E
Scanning Speed High, Medium, Low (Max. 6000nm/min)
Cuvette Holder 10mm 4-cell holder
Output Cloud connectivity, Email, SMS, Bluetooth, WiFi
Display 11-inch Touch Screen
Power AC 110 to 220V/50 to 60Hz
Package Dimension 810×660×390 mm
Gross Weight 27 kg

Features

11-inch touch screen

Low stray light and noise

Silicon photodiode detector

Rapid scanning speed

Multiple working modes

Real-time data sending

Adopts 3.0 Bluetooth protocol

Cloud and wireless output options

Standard Accessories

Model No.

Accessory

LMUS-D100-G1

10mm Glass cuvette

LMUS-D100-Q1

10mm Quartz cuvette

LMUS-D100-P1

Power cord

Optional Accessories

Model No.

Accessory

LMUS-D100-G2

5mm Glass Cuvette
LMUS-D100-G310mm Glass Cuvette
LMUS-D100-G420mm Glass Cuvette
LMUS-D100-G530mm Glass Cuvette
LMUS-D100-G650mm Glass Cuvette
LMUS-D100-G7100mm Glass Cuvette

LMUS-D100-Q2

5mm Quartz Cuvette
LMUS-D100-Q310mm Quartz Cuvette
LMUS-D100-Q420mm Quartz Cuvette
LMUS-D100-Q530mm Quartz Cuvette
LMUS-D100-Q650mm Quartz Cuvette
LMUS-D100-Q7100mm Quartz Cuvette

LMUS-D100-QM1

0.35/0.7/1.0/1.4/1.7ml Quartz Micro-Cuvette

LMUS-D100-QM2

50/100/200µl Quartz Micro-Cuvette

LMUS-D100-GM1

0.35/0.7/1.0/1.4/1.7ml Glass Micro-Cuvette

LMUS-D100-GM2

50/100/200µl Glass Micro-Cuvette

LMUS-D100-H1

10mm Manual 4-cell Holder

LMUS-D100-H2

50mm Manual 4-cell Holder

LMUS-D100-H3

100mm Manual 4-cell Holder

LMUS-D100-S1

10mm Single-hole Cell Holder

LMUS-D100-SH1

50mm Single-hole long-path cell Holder

LMUS-D100-SH2

100mm Single-hole long-path cell Holder

LMUS-D100-A1

Auto 8-cell/4-cell/5-cell Holder

LMUS-D100-F1

Film Holder

LMUS-D100-R1

Tube Rack

LMUS-D100-M1

Adjustable XY Micro Cell Holder

LMUS-D100-PS1

Peltier/Sipper System

LMUS-D100-I1

Integrating Sphere

LMUS-D100-T1

Tungsten Lamp

LMUS-D100-X1

Xenon Lamp

LMUS-D100-TP1

Thermal Printer 

LMUS-D100-TP2

Built-in Thermal Printer

LMUS-D100-PC1

PC software

LMUS-D100-CF1

21 CFR software

Applications

Split Beam UV-Vis Spectrophotometer LMUS-D100 is suitable for measurement of absorbance and concentration in pharmaceutical quality control, biological research, environmental analysis, food testing, and scientific laboratories.

Frequently Asked Questions

1 What is a Split Beam UV-Vis Spectrophotometer?

A Split Beam UV-Vis Spectrophotometer is an apparatus designed to compare the intensity of light passing through a sample with a reference, improving accuracy and stability in UV-Visible spectroscopy.

2 How does Split Beam UV-Vis Spectrophotometer work?

It works by splitting light into two beams, one passing through the sample and the other through a reference, to correct for fluctuations and improve accuracy.

3 Can Split Beam UV-Vis Spectrophotometer be used for quantitative analysis?

It is widely used for quantitative analysis due to its ability to deliver accurate and reproducible measurements by continuously comparing the sample and reference beams.

4 What kind of maintenance is required by Split Beam UV-Vis Spectrophotometer?

Routine maintenance includes regular cleaning of optical components, checking of lamp condition, and occasional recalibration to maintain performance.

5 What types of samples can be analyzed using Split Beam UV- Vis Spectrophotometer?

It is used for analyzing liquids, solids, powders, and thin films in various scientific and industrial fields.

6 How is Split Beam UV-Vis Spectrophotometer used in pharmaceutical analysis?

It is widely used for quantifying active pharmaceutical ingredients (APIs), checking drug purity, dissolution testing, and studying drug stability.

7 Can Split Beam UV-Vis Spectrophotometer monitor heavy metals in water?

It can detect heavy metals like chromium, iron, or lead by forming colored complexes with specific reagents and measuring absorbance.

8 How is Split Beam UV-Vis Spectrophotometer applied in chemical industries?

It is used for monitoring chemical reaction kinetics, analyzing dye concentrations, measuring metal ion content, and quality control of chemical formulations.

9 How is Split Beam UV-Vis Spectrophotometer used in materials science?

It is used to analyze optical properties of coatings, determine transmittance and absorbance of materials, and study semiconductor band gaps.

10 Can Split Beam UV-Vis Spectrophotometer be used for protein and DNA analysis?

It is commonly used to quantify nucleic acids at 260 nm and proteins at 280 nm, and to assess sample purity using A260/A280 ratio.