Features :
Large LCD display with touch screen or mouse 8-channel optical fibre system i.e. 5 sec reading for the whole plate Bio chromatic measurement and calculations Storage of 100 test protocol and 10000 sample results Multi-assay feature – up to 12 different assay on a plate Adjustable shaking time and speed Random positive and negative control setting Powerful QC functions and programs Auto self-check when power on
Specifications :
| Plate type | 96 well plate |
| Display | 6 ” LCD |
| Standard Wavelength | 405, 450, 492, 630 nm |
| Wavelength accuracy | ± 1 nm |
| Resolution | 0.001 Abs |
| Reading speed (Single wavelength) | 5 secs for 96 well plate |
| Photometric Range | 0.000 – 4.000 Abs |
| Accuracy | ± 1 % or ± 0.005 A |
| Optic system | 8-channel optic fibre system |
| Band width | < 8 nm |
| Calculation mode | ABS, %ABS, Cut-off, Curve, Linear, Log, Exponent, Power regression |
| Memory | Upto 10,000 tests results and 100 test protocol |
| Interface | RS - 232, USB, SD Card |
| Light source | Halogen lamp |
| Input | Touch screen, mouse or external keyboard |
| Power supply | AC 100 V – 220 V ± 10 %, 50-60 Hz |
| Dimensions | 460 x 330 x 200 mm |
| Weight | 8 Kg |
Selection Guide :
Microplate Reader LMRW-A101 is a versatile analytical tool commonly used in a wide range of research and diagnostic applications. It is particularly suited for studies involving enzyme kinetics, protein quantification (e.g., BCA, Bradford, Lowry assays), ELISA, and cell viability assays. Its design supports high-throughput analysis, making it valuable in pharmaceutical screening, clinical diagnostics, and biotechnology research. The reader is compatible with 96-well and 48-well microplates, ensuring flexibility for different assay formats. It typically provides a broad wavelength range, which supports diverse colorimetric assays. Customers should consider this model if their research demands accurate absorbance readings, consistent performance, and regulatory compliance, as this reader generally conforms to CE certification standards. Its user-friendly interface, reliable optical system, and compact footprint further enhance its usability in laboratory environments focused on precision and reproducibility.