Get Quote
Contact us now to receive a detailed quotation.
Overview
Labmate Pulse Vacuum Sterilizer LMPVS-B202 offers a 360 L chamber with a vertical sliding door, sized for high-volume sterile processing. A 30 kW steam generator brings cycles up to 139 °C, while vacuum and pressure pulsing down to -96 kPa removes trapped air. Vacuum drying leaves instruments below 0.1% moisture. This large autoclave sterilizer processes full loads of instruments, textiles, glassware and culture media in a single cycle.
Features :
Chamber Access
Vertical sliding door
Load Range
Porous, hard and non-sealed liquid loads
Cycle Programs
Built-in cycles for gowns, lumen instruments, rubber and liquids
Automation
Fully automatic PLC control with status display
Documentation
Built-in printer for cycle records
Air Removal
Negative, transpulse and positive pressure pulsing
Steam Supply
Integrated steam generator system
Load Drying
Vacuum flash drying for fabrics and instruments
Thermal Regulation
Uniform chamber temperature regulation
Safety Valves
Separate chamber and jacket relief valves
Specifications :
| Chamber Volume | 360 L |
| Chamber Size (W × H × D) | 610 × 610 × 1000 mm |
| Maximum Operating Temperature | 139 °C |
| Design Temperature | 150 °C |
| Temperature Control Accuracy | ±1 °C |
| Temperature Uniformity | ±1 °C |
| Temperature Resolution | 0.1 °C |
| Pressure Resolution | 0.1 kPa |
| Maximum Operating Pressure | 0.25 MPa |
| Chamber Design Pressure | -0.1 to 0.3 MPa |
| Jacket Design Pressure | 0.3 MPa |
| Steam Generator Design Pressure | 1.0 MPa |
| Maximum Vacuum | -96 kPa |
| Negative Pressure Pulsing | -80 to 0 kPa |
| Transpulse (Across-Pressure Pulsing) | -80 to 80 kPa |
| Positive Pressure Pulses | 0 to 9 times |
| Drying Performance | Fabric moisture < 0.6%, instrument moisture < 0.1% |
| Chamber Relief Valve | 0.28 MPa |
| Jacket Relief Valve | 0.28 MPa |
| Steam Consumption | 20 kg/cycle |
| Water Consumption | 160 kg/cycle |
| Power Supply | 380 V, 50 Hz |
| Power | 2 kW control + 30 kW steam generator |
| External Dimensions (L × W × H) | 1350 × 1350 × 1850 mm |
| Packaging Dimensions (L × W × H) | 1500 × 1500 × 2050 mm |
| Net Weight | 870 kg |
| Gross Weight | 945 kg |
Applications :
Applications: Hospital Sterile Supply Units | Biopharmaceutical Production | Microbiology Laboratories
Functions: Textile and gown sterilization | Glassware and container sterilization | Culture media sterilization