Get Quote
Contact us now to receive a detailed quotation.
Overview
Photometer LMPTR-100 features a range of wavelengths, each equipped with an interference filter ranging from 5 nm or 6 nm in bandwidth. It operates within a temperature span of 5°C to 40°C and a humidity range of 30% to 90%. Our photometer offers 120 pre-programmed testing methods for versatile water testing needs. It is integrated with a graphic display providing an intuitive interface and interpreting the measurements.
Features :
One-time-Zero function
Auto-off capability
IP 68 protection glass
Incorporated with a beeper
Supports large data storage
Complies with CE standards
Specifications :
| Wavelength Range | 430 nm IF Δλ = 5 nm 530 nm IF Δλ = 5 nm 560 nm IF Δλ = 5 nm 580 nm IF Δλ = 5 nm 610 nm IF Δλ = 6 nm 660 nm IF Δλ = 5 nm |
| Optics | LEDs, interference filters (IF), and photo sensors in transparent sample chamber |
| Wavelength Accuracy | ± 1 nm |
| Photometric Accuracy | 2 % FS (T = 20 °C to 25 °C) |
| Suitable Vials | Multi vial 10 mm Round Cuvettes 13 mm Round Cuvettes 16 mm Round Cuvettes 24 mm |
| Display | Graphic-display |
| Interfaces | Infrared |
| Operating Conditions | 5 to 40 °C at maximum 30 to 90 % RH |
| Operation | Acid and solvent-resistant, touch-sensitive keypad with audible feedback via integrated beeper |
| Auto – OFF | Yes |
| Data Storage | 1000 data sets |
| Protection Class | IP 68 |
| Languages User Interface | German, English, French, Spanish, Italian, Portuguese, Polish, Indonesian |
| Compliance | CE |
| Battery Lifetime | 26 hours |
| Power Supply | 4 batteries (Mignon AA/LR6) |
| Dimensions | 95 × 45 × 210 nm |
| Weight | 450g |
Standard Accessories :
Glass bottles with a volume of 1.10ml×3 2.10mm diameter cuvette×1 Adapter×1
Optional Accessories :
Digester
Applications :
Photometer is widely used in water quality analysis, biomedical research, food and beverage industry, industrial process control, and environmental monitoring.