Get Quote
Contact us now to receive a detailed quotation.
Overview
Oil Tan Delta Tester LMODT-A100 has capacitance measuring range of 5 to 200 pF. It features non-contact mid-frequency induction heating for uniform, fast, and stable temperature control. Equipped with solenoid drain switch, it allows complete draining and rinsing without removing oil cup. It integrates real-time clock support to store, display, and print date-stamped test results automatically. Our Tan Delta Tester checks dielectric loss and resistivity of insulating oils in labs and substations.
Features :
Backlit LCD display
Three-electrode cup
Overload protection
Sensor failure warning
Thermal safety relay
Auto result storage
Result printing
Specifications :
| Capacitance Measuring Range | 5 to 200 pF |
| Relative Permittivity Range | 1.000 to 30.000 |
| Dielectric Loss Factor Range | 0.00001 to 100 |
| DC Resistivity Range | 2.5 MΩm to 20 TΩm |
| Capacitance Accuracy | ± (1% of reading + 0.5pF) |
| Relative Permittivity Accuracy | ± 1% of reading |
| Dielectric Loss Factor Accuracy | ± (1% reading +0.0001) |
| DC Resistivity Accuracy | ± 10% of reading |
| Capacitance Resolution | 0.01pF |
| Relative Permittivity Resolution | 0.001 |
| Dielectric Loss Factor Resolution | 0.00001 |
| Temperature Range | 0 to 125 ℃ |
| Temperature Measurement Error | ± 0.5 ℃ |
| AC Test Voltage | 500 to 2200V continuously adjustable frequency 50Hz |
| DC Test Voltage | 0 to 500V continuously adjustable |
| Operating Temperature | 0 to 40 ℃ |
| Operating Humidity | Less than 75% RH |
| Power Consumption | 100W |
| Power Supply | AC 220V ± 10%, 50Hz / 60Hz ± 1% |
| Dimension | 500×360×420 mm |
| Weight | 22Kg |
Applications :
Oil Tan Delta Tester LMODT-A100 measures dielectric loss and DC resistivity of insulating oils and liquid dielectrics in power systems, laboratories, transformer maintenance, and quality control environments.
Selection Guide :
Oil Tan Delta Tester LMODT-A100 is a reliable instrument designed for evaluating the insulation performance of transformer oils through accurate measurement of tan delta, capacitance, and DC resistivity. Based on research studies, it is widely used in applications such as transformer maintenance, quality assurance of insulating oils, and high-voltage equipment diagnostics in power utilities, electrical labs, and R&D facilities. It is particularly effective for experiments involving dielectric loss factor analysis and dissipation factor evaluation, helping to detect aging, moisture, and contamination in insulating fluids. The tester complies with international standards, such as IEC 60247, ensuring credibility and accuracy in the results. Customers can select this model based on key parameters, including its wide capacitance measuring range, high-resolution tan delta accuracy, DC resistivity range, and stable performance under various temperature conditions. Its robust design and certified performance make it perfect for professionals requiring advanced insulation diagnostics and compliance-based testing.