Gas Chromatography LMCG-101 has temperature range of 7°C to 450°C. It ensures stable thermal control and efficient column performance. Equipped with an advanced detector that delivers consistent results. Integrated with an intuitive touchscreen that allows seamless operation. Our Gas Chromatography supports analytical settings in pharmaceuticals, petrochemicals, and environmental testing.
Features :
Computer-controlled system
Multistage temperature control
Real-time status monitoring
Rapid heat-up and cool-down
Dual injector operation
Independent signal output
Specifications :
| Temperature Range | 7ºC to 420ºC |
| Temperature Accuracy | ±0.02ºC |
| Cooling Time | 350°C to 100°C ≤ 3 min 400°C to 50°C in 8-10 min at 25°C ambient |
| Programmed Temperature Range | 0.1℃ to 80℃/min |
| Program Ramps | 20 Ramps |
| FID Detection Time | ≤8×10-12 g/s (n-hexadecane) |
| FID Baseline Drift | ≤1.5×10-13 A/h |
| FID Baseline Noise | ≤3×10-14 A |
| FID Linearity | ≥106 |
| TCD Maximum Operating Temperature | 400ºC |
| TCD Temperature Accuracy | ±0.1ºC |
| TCD Detector Sensitivity | ≥ 10000 mV × mL / mg (n-Hexadecane) |
| TCD Baseline Noise | ≤20uv |
| TCD Baseline Drift | ≤100uv/30 min |
| TCD Linear Dynamic Range | ≥104 |
| FPD Detection | ≤8×10‐13 g/s(P); ≤8×10‐11 g/s(S) |
| FPD Baseline Drift | ≤2×10‐11 A/30 min |
| FPD Baseline Noise | ≤5×10‐12 A |
| NPD Detection Limit | ≤5×10-12 g/s(N) (Azobenzene) ≤5×10-13 g/s(P) (Malathion) |
| ECD Detection Limit | ≤3×10-14 g/ml |
| ECD Baseline Drift | ≤ 100 μV/30 min |
| ECD Baseline Noise | ≤40μV |
| Internal Dimension (W×D×H) | 284×280×241 mm |
| External Dimension (W×D×H) | 340×345×281 mm |
Standard Accessories :
M6 Software
Optional Accessories :
Model No. | Accessory | Quantity | Unit |
LMCG-101-GC1 | Split/Non-split capillary injector | 1 | Each |
LMCG-101-GC2 | Flame Ionization Detector (FID) | 1 | Each |
LMCG-101-GC3 | Thermal Conductivity Detector (TCD) | 1 | Each |
LMCG-101-GC4 | Flame Photometric Detector (FPD) | 1 | Each |
LMCG-101-GC5 | Electron Capture Detector (ECD) | 1 | Each |
LMCG-101-GC6 | Nitrogen Phosphorus Detector (NPD) | 1 | Each |
LMCG-101-GC7 | Autosampler 22 places | 1 | Each |
LMCG-101-GC8 | Autosampler 156 places | 1 | Each |
Applications :
Gas Chromatography LMCG-101 is employed in pharmaceuticals, petrochemicals, food safety, and environmental analysis. It ensures reliable compound separation and detection in complex mixtures.
Selection Guide :
Gas Chromatography LMCG-101 is designed for the separation and analysis of volatile and semi-volatile compounds. It is widely used in applications such as environmental monitoring, pharmaceutical quality control, food and beverage testing, petrochemical analysis, and forensic investigations. Based on research studies, it is particularly suitable for quantitative and qualitative analysis in both routine and complex experiments. This system provides a temperature control range up to 450°C, high-accuracy sample injection, and advanced electronic pressure control (EPC) for stable performance. It meets international certifications like ISO and CE, ensuring reliability and compliance with laboratory standards. Customers looking for a robust system for research, QA/QC labs, or regulatory testing can confidently choose this model based on its proven performance, certified quality, and versatile application support.