Fluorescence Spectrophotometers measure the light a sample emits after absorbing light at a shorter wavelength, detecting many compounds at far lower concentrations than absorbance methods. Dual monochromators select excitation and emission independently across 200 to 900 nm, with slit widths of 1 to 20 nm to trade sensitivity against resolution. Wavelength accuracy holds within 0.1 to 1 nm and repeatability within 0.2 to 0.5 nm, while signal to noise reaches 150:1 on the water Raman peak and up to 1000:1 RMS on high-end models. Typical uses include DNA and protein quantification, enzyme kinetics, biomarker assays, drug formulation QC, pollutant screening and quantum dot research.
Fluorescence Spectrophotometer LMFSP-A100 Catalog
| Wavelength Range | 250 to 700nm |
| Wavelength Accuracy | ±1 nm |
| Wavelength Repeatability | Less than or equal to 0.5 nm |
| Detection Sensitivity | Signal-to-noise ratio greater than or equal to 150 |
| Measurement Source | Raman peak of water in a 1 cm quartz cuvette |
Fluorescence Spectrophotometer LMFSP-A101 Catalog
| Excitation Wavelength | 200nm to 900nm |
| Emission Wavelength | 200nm to 900nm |
| Excitation Slit | 10nm |
| Emission Slit | 10nm |
| Wavelength Accuracy | ±0.1nm |
Fluorescence Spectrophotometer LMFSP-A102 Catalog
| Wavelength Range | 200 to 900 nm |
| Bandwidth | 1nm/2nm/5nm/10nm/20nm |
| Wavelength Accuracy | ±0.4nm |
| Wavelength Repeatability | Less than or equal to 0.2nm |
| Signal-To-Noise Ratio | Greater than or equal to 350(P-P) or Greater than or equal to 1000(RMS) |