Get Quote
Contact us now to receive a detailed quotation.
Overview
Bilirubin Meter LMBIM-A100 comes with a measuring range of 0-30 mg/dL and offers a broad spectrum of analysis. It requires a sample volume of only 40-45 µL, filling 2/3 to 3/4 of the total capillary volume. Equipped with long-life white LED light source which ensures longevity. Incorporated with built-in thermal printer provides quick, hassle-free printouts. Our Bilirubin Meter features a disposable capillary for accurate sample collection.
Features :
High humidity tolerance
Atmospheric pressure adaptability
RS232 connectivity
Memory storage capacity
Bi-chromatic photometer
Specifications :
| Measuring Range | 0-30 mg/dL |
| Sample Volume | 40-45 µL (2/3 to 3/4 of the total capillary volume) |
| Environment Temperature | 10°C to 30°C |
| Relative Humidity | ≤95% |
| Atmospheric Pressure | 860 hpa to 1060 hpa |
| Warm Up | ≥5 min |
| Capillary | 75mm/60µL disposable capillary, diameter 1.3mm-1.4mm (Hirschmann Laborgerate product REF.9100260 recommended) |
| Measuring Principle | Direct photometry |
| Built-in Printer | 58mm thermal printer |
| Light Source | Long-life time white LED |
| Connection Port | USB port |
| Photometer | Bi-chromatic, 450nm (M) and 578nm (S) filter |
| Display | 16x2 ASC code LCD with backlight |
| Interface | Standard RS232 |
| Power Supply | AC 100-240V, 50/60Hz |
| Dimensions(L× W× H) | 250 × 190× 85 mm |
| Packaging Dimensions (L× W× H) | 320 × 250 × 170 mm |
| Weight | 2 kg (N.W.), 3 kg (G.W.) |
Standard Accessories :
Main Machine × 1 piece User's Manual × 1 piece Power Adapter (12V 3A) × 1 piece Thermal print Paper Roll × 2 pieces RS232 Cable × 1 piece Dust Cover × 1 piece Standard capillary × 100 pieces The seal plate × 1 piece
Applications :
Bilirubin Meter LMBIM-A100 is essential in clinical laboratories, neonatal care units, and healthcare facilities for the accurate measurement of bilirubin levels in blood samples.