Benchtop Turbidity Meter LMBTM-401 is equipped with a 4.5 inches TFT display. TSS conversion factor ensures precise measurement of the total suspended solids. 200 sets of data are stored and can be transferred to computer by USB communication interphase. Password protection inhibits the unauthorized calibration and settings. It delivers highly precise results with better numerical calculations to meet GLP standards. Continuous measurement mode is used for indexing and matching the sample vials.
Turbidity Range | 0 to 2000 NTU, 0 to 500 EBC, 0 to 9999 ASBC |
Turbidity Resolution | 0.01 (0 to 99 NTU), 0.1 (100 to 999 NTU), 1 (1000 to 2000 NTU) |
Turbidity Accuracy | ±2% of reading (0 to 500 NTU), ±3% of reading (501 to 2000 NTU) |
Calibration Points | 2 to 7 points |
Calibration Standards | 0.02, 10, 200, 500, 1000, 1500, 2000 NTU |
TSS Range | Depending on the TSS conversion factor |
TSS Resolution | 0.1, 1 |
TSS Accuracy | ±3% of reading |
Light Source | Infrared-emitting diode (850 nm wavelength) |
Detector | Silicon photodiode |
Stray Light | <0.02 NTU |
Sample Volume | 30 ml |
Memory | Stores up to 200 data sets |
Output | USB communication interface |
Display | 4.5" TFT LCD |
Power | DC12V, using AC adapter, 220V/50 Hz |
Dimensions(L×W×H) | 250×177×96 mm |
Weight | 1.2 Kg |
Widely used in swimming pool, water recycling units, water treatment plants, research centers and laboratories of pharmacy, food industry and beverage industry