Get Quote
Contact us now to receive a detailed quotation.
Overview
Automatic Coagulation Analyzer LMCA-B100 measures PT, APTT, TT, fibrinogen and D-dimer by optical coagulation and immunoassay methods across four test channels. A 40-cuvette module with 10 sample and 10 reagent positions supports continuous 24-hour workflows, while automatic probe cleaning, re-testing of abnormal results and error alerts reduce manual handling. Detector and sample disk temperature holds at 37 °C ±1 °C for consistent clotting times. See more Automatic Coagulation Analyzer options from Labmate for laboratory sample processing.
Features :
Specifications :
| Method | Optical coagulation assay, and immunoassay |
| Test Channels | 4 |
| Test Methods | Automatic or manual test mode |
| Sample Position | 10 |
| Reagent Container Position | 10 |
| Cuvette | 40 |
| Test Timing Range | 0 to 300 s |
| Test Speed | PT - 300 Tests/h, APTT - 250 Tests/h, TT - 300 Tests/h, FIB – 300 Tests/h |
| Reagent Volume | 5 to 200 μL |
| Sample Volume | 5 to 100 μL |
| Preheating Time | >30 mins (after start-up) |
| Constant Temperature | Detector: 37℃ ± 1℃, Reagent and Sample disk: 37℃ ± 1℃ |
| Test Items | Prothrombin time (PT), activated partial thromboplastin time (APTT), thrombin time (TT), fibrinogen (FIB), coagulation factors II-XII, D-dimer, reptilase time (RT), lupus anticoagulant (LA), heparin (HEP), protein C (PC), etc. |
| Temperature Control | Test module and preheating system: 37℃ ± 1℃ (constant temperature control)Reagent cooling temperature control: >16℃ |
| Repeatability | CVPT ≤ 2% CVAPTT ≤ 2.5% CVTT ≤ 3% CVFIB≤5% CV D-Dimer ≤ 6% |
| Relative Bias | FIB relative bias ≤ ±10.0%, D-dimer relative bias ≤ ±10.0% |
| Linearity | FIB linearity r ≥ 0.975 D-dimer linearity r ≥ 0.975 |
| Sample Carryover Rate | FIB carryover rate ≤10% |
| Continuous Working Time | ≥ 24 h |
| Display | LCD display or touchscreen |
| Storage | More than 50 million pieces of data |
| Power Supply | AC 220 ± 22 V AC, 50 ± 1 Hz |
| Power | ≤ 320 VA |
| Packaging Dimension (W×D×H) | 820 × 730 × 740 mm |
| Net Weight | 39 kg |
| Gross Weight | 75 kg |
Selection Guide :
Automatic Coagulation Analyzer LMCA-B100 is designed for accurate and efficient analysis in clinical, research, and experimental applications. It supports optical coagulation assay and immunoassay methods with 4 test channels, enabling multiple tests simultaneously for improved workflow. Users can choose between automatic and manual test modes, providing operational flexibility based on workload and protocols. With 10 sample positions, it ensures streamlined sample handling for routine testing or study applications. Customers should select this model for its accurate performance, certified reliability, and suitability across research studies, laboratory experiments, and diagnostic applications.