Get Quote

Contact us now to receive a detailed quotation.

Need quick help? Chat with us on WhatsApp:

WhatsApp QR Code

OR

Overview

Automatic Coagulation Analyzer LMCA-B100 measures PT, APTT, TT, fibrinogen and D-dimer by optical coagulation and immunoassay methods across four test channels. A 40-cuvette module with 10 sample and 10 reagent positions supports continuous 24-hour workflows, while automatic probe cleaning, re-testing of abnormal results and error alerts reduce manual handling. Detector and sample disk temperature holds at 37 °C ±1 °C for consistent clotting times. See more Automatic Coagulation Analyzer options from Labmate for laboratory sample processing.

Features :

  • Compact benchtop design with LCD display and remote control.
  • Continuous 24-hour operation.
  • Self-testing with editable calibration.
  • User-friendly interface with consistent performance.

Specifications :

Method Optical coagulation assay, and immunoassay
Test Channels 4
Test Methods Automatic or manual test mode
Sample Position 10
Reagent Container Position 10
Cuvette 40
Test Timing Range 0 to 300 s
Test Speed PT - 300 Tests/h,
APTT - 250 Tests/h,
TT - 300 Tests/h,
FIB – 300 Tests/h
Reagent Volume 5 to 200 μL
Sample Volume 5 to 100 μL
Preheating Time >30 mins (after start-up)
Constant Temperature Detector: 37℃ ± 1℃,
Reagent and Sample disk: 37℃ ± 1℃
Test Items Prothrombin time (PT), activated partial thromboplastin time (APTT), thrombin time (TT), fibrinogen (FIB), coagulation factors II-XII, D-dimer, reptilase time (RT), lupus anticoagulant (LA), heparin (HEP), protein C (PC), etc.
Temperature Control Test module and preheating system: 37℃ ± 1℃ (constant temperature control)Reagent cooling temperature control: >16℃
Repeatability CVPT ≤ 2%
CVAPTT ≤ 2.5%
CVTT ≤ 3%
CVFIB≤5%
CV D-Dimer ≤ 6%
Relative Bias FIB relative bias ≤ ±10.0%,
D-dimer relative bias ≤ ±10.0%
Linearity FIB linearity r ≥ 0.975
D-dimer linearity r ≥ 0.975
Sample Carryover Rate FIB carryover rate ≤10%
Continuous Working Time ≥ 24 h
Display LCD display or touchscreen
Storage More than 50 million pieces of data
Power Supply AC 220 ± 22 V AC, 50 ± 1 Hz
Power ≤ 320 VA
Packaging Dimension (W×D×H) 820 × 730 × 740 mm
Net Weight 39 kg
Gross Weight 75 kg

Selection Guide :

Automatic Coagulation Analyzer LMCA-B100 is designed for accurate and efficient analysis in clinical, research, and experimental applications. It supports optical coagulation assay and immunoassay methods with 4 test channels, enabling multiple tests simultaneously for improved workflow. Users can choose between automatic and manual test modes, providing operational flexibility based on workload and protocols. With 10 sample positions, it ensures streamlined sample handling for routine testing or study applications. Customers should select this model for its accurate performance, certified reliability, and suitability across research studies, laboratory experiments, and diagnostic applications.

Labmate Scientific LLC