Get Quote
Contact us now to receive a detailed quotation.
Overview
Labmate ATP Luminometer LMAHM-A100 detects ATP from surface swab samples through bioluminescence, with a 0 to 999,999 RLU detection range and results within 15 seconds. Detection sensitivity down to 1 × 10⁻¹⁸ mol ATP identifies low residual contamination, while settable upper and lower thresholds give automatic pass/fail evaluation. As a handheld hygiene monitoring system, the luminometer supports on-site sanitation verification and contamination control.
Features :
Detection System
Bioluminescence ATP detection
Test Capability
ATP, ALP and milk allergen testing
Measurement Performance
High-sensitivity RLU measurement
Result Evaluation
Automatic pass/fail judgement
Calibration System
Self-calibrating light source
Data Management
PC data transfer via USB
Specifications :
| Detection Principle | Firefly luciferase–luciferin bioluminescence |
| Detection Sensitivity | 1 × 10⁻¹⁵ to 1 × 10⁻¹⁸ mol ATP |
| Detection Range | 0–999,999 RLU |
| Detection Mode | RLU |
| Testing Time | 15 s |
| Deviation | ±5% or ±5 RLU |
| ATP Recovery Rate | 90–110% |
| Pass/Fail Evaluation | Settable upper and lower limits with automatic pass/fail judgement |
| Calibration | Built-in self-calibrating light source |
| Self-Check | Automatic self-check 30 s after start-up |
| Display | 3.5-inch graphic touch screen |
| Processor | 32-bit data processing chip |
| Data Transfer | Mini USB interface for uploading results to a PC |
| Software | Data analysis software on USB drive |
| Battery | 3.7 V, 3000 mAh lithium battery |
| Standby Time | 6 months (at 20°C) |
| Operating Temperature | 5–40°C |
| Operating Humidity | 20–85% |
| Product Dimensions (W × H × D) | 188 × 77 × 37 mm |
| Packaging Dimensions | 400 x 300 x 200 cm |
| Net weight | 400 g |
| Gross Weight | 3 kg |
Standard Accessories :
Applications :
Applications: Food Processing | Healthcare Facilities | Water Treatment
Functions: Surface hygiene verification | Cleaning validation | Contamination screening