Atomic Emission Spectrophotometer LMAES-A100 operates within wavelength range of 165 nm to 950 nm. It ensures reliable elemental analysis across various sample types. Equipped with advanced CCD detector technology which provides enhanced signal detection and resolution. Integrated with an intuitive touchscreen interface that simplifies operation and real-time data monitoring. Our Spectrophotometer is ideal for elemental analysis in environmental, pharmaceutical, and metallurgical industries.
Features :
Automatic protection mechanisms
Flame monitoring system
RF reverse power alert
Fast sample injection
Advanced spectral analysis
Durable and reliable build
Specifications :
| Spectral Range | 165 nm to 950 nm |
| Focal Length | 400 mm |
| Grating | Echelle Grating |
| Prism | Ultra-pure SiO2 material |
| Stray Light | Equivalent background concentration of 10000μg/ml Ca solution |
| Wavelength | 189.042 nm |
| Stray Light Value | < 2μg/mL |
| Element Detection Limit | 1to 10 ppb |
| Number of Pixels | 1024×1024 |
| Pixel Area | 24µm x 24µm |
| Stabilization Time | <3 minutes |
| Oscillation Frequency | 27.12 MHz |
| Electromagnetic Field Leakage Radiation Intensity | <0.5V/m |
| Torch Coil | 3 turns |
| Torch | Three concentric quartz torch |
| Atomizer | Concentric or Parallel channel |
| Spray Chamber | Cyclonic spray type |
| Peristaltic Pump | 4 Channels, 12 rollers |
| Argon Gas Consumption | 8 to 18 L/min |
| Cooling Gas Consumption | 0.00 to 20 L/min |
| Auxiliary Gas Consumption | 0 to 2 L/min |
| Carrier Gas Consumption | 0 to 2 L/min |
| Accuracy | 0.01 L/min |
| Solid Content | 0.001 to 70 % |
| Liquid Content | 0.01 to 1000 ppm |
| Repeatability | RSD <1% |
| Stability | <1.5% |
| Torch Outer Diameter | 20mm |
| Atomizer Outer Diameter | 6mm |
| Working Temperature | 15 to 30℃ |
| Working Humidity | ≤70% |
| Atmospheric Pressure | 86 to 106 kPa |
| Grating Size | 100mm x 50mm |
| Power Consumption | 700W to 1600W |
| Power Supply | 220 ±22V AC, 50 ±1 Hz |
Standard Accessories :
Accessory | Unit |
ICP-6810-A2 Host | 1 |
ICP Analysis Software | 1 |
U Tray | 1 |
Parvicostellae | 1 |
Plastic Cask | 1 |
Argon Pressure Relief Valve | 2 |
30M PU Flexible Pipe | 1 |
PU Flexible Pipe | 5 |
PVC Clear Tube Hose | 1 |
Plug Power Cord | 1 |
Ground Copper Woven Wire | 15 |
Brass Pipe | 1 |
Filter Tank Filter Element | 4 |
Torch Tube | 2 |
Double Cylinder Fog Chamber | 1 |
Atomization Device | 1 |
Capillary into the Sample Tube | 0.5 |
Peristaltic Pump Head Hose | 3 |
Tube Fuse | 2 |
Tube Fuse | 2 |
Conoid Fog Chamber Cover | 1 |
4 M Silicone Tube | 1 |
Hose Transfer Connector | 4 |
Hose Transfer Connector | 4 |
Ball Interface Clip | 1 |
Type O Sealing Ring | 1 |
Type O Sealing Ring | 2 |
Quick Fitting | 4 |
PTFE Tape | 1 |
Ribbon | 30 |
Torch Pipe Tooling | 1 |
High Frequency Coil Assembly | 1 |
Main Engine | 1 |
Indicator | 1 |
Printer | 1 |
Coolant Box | 1 |
Stabilized Voltage Supply | 1 |
Installed Parts | 1 |
Optional Accessories :
Auto Sampler
Organic Injection System
Applications :
Atomic Emission Spectrophotometer LMAES-A100 is employed for trace element analysis in wide range of fields including environmental protection, petroleum, metallurgy, and clinical medicine. It is ideal for analyzing samples such as geological materials, food, biological samples, and more.