Biological Safety Cabinet Class II Type B2

Biological Safety Cabinets Class II Type B2 protect the operator, the sample and the room by drawing air in at the front, sending HEPA filtered air down over the work surface and ducting 100% of the exhaust out of the building, with no air recirculated. Total exhaust makes Type B2 the choice when work combines infectious agents with volatile toxic chemicals or radionuclides. Inflow holds at 0.53 m/s and downflow at 0.33 m/s, cabinets come in 43 to 73 inch widths with exhaust volumes of 360 to 1320 m3 per hour, and HEPA filters capture 99.999% of particles at 0.3 um. Typical uses include TB culture, bloodborne pathogen work, toxic drug handling and animal research.

 Compare

Biological Safety Cabinet Class Ⅱ Type B2 LMBY-A100

  • Cabinet Size: 43 Inches
  • Inflow Velocity: 0.53 ± 0.025 m/s
  • Down flow Velocity: 0.33 ± 0.025 m/s
  • Exhaust Volume: 360 m³/h
  • Pre-filter: Polyester Fiber Filter

$ 6275 / Each of 1

 Compare

Biological Safety Cabinet Class Ⅱ Type B2 LMBY-A101

  • Cabinet Size: 51 Inches
  • Inflow Velocity: 0.53 ± 0.025 m/s
  • Down flow Velocity: 0.33 ± 0.025 m/s
  • Exhaust Volume: 500 m³/h
  • Pre-filter: Polyester Fiber Filter

$ 6737.5 / Each of 1

 Compare
Biological Safety Cabinet Class Ⅱ Type B2 LMBY-A102
  • Cabinet Size: 59 Inches
  • Inflow Velocity: 0.53 ± 0.025 m/s
  • Down flow Velocity: 0.33 ± 0.025 m/s
  • Exhaust Volume: 880 m³/h
  • Pre-filter: Polyester Fiber Filter

$ 7387.5 / Each of 1

 Compare
Biological Safety Cabinet Class Ⅱ Type B2 LMBY-A103
  • Cabinet Size: 73 Inches
  • Inflow Velocity: 0.53 ± 0.025 m/s
  • Down flow Velocity: 0.33 ± 0.025 m/s
  • Exhaust Volume: 1270 m³/h
  • Pre-filter: Polyester Fiber Filter

$ 9462.5 / Each of 1

 Compare

Biological Safety Cabinet Class Ⅱ Type B2 LMBY-B100

  • Maximum Opening: 550 mm (22 inch)
  • Tested Opening: Safety height 200 mm (8 inch)
  • Exhaust airflow volume: 100%: 1320 m³/h (776 cfm)
  • Front Window: Motorized. Two-layer laminated toughened glass ≥ 5 mm. Anti UV
  • HEPA Filter: Two, 99.999% efficiency at 0.3 µm, filter life indicator

$ 7525 / Pack of 1

Frequently Asked Questions

1 What is a Biological Safety Cabinet Class II Type B2?

A Biological Safety Cabinet Class II Type B2 is an apparatus designed to provide personnel, product and environmental protection by exhausting 100% of filtered air outside the facility.

2 What type of filters does Biological Safety Cabinet Class II Type B2 use?

It uses HEPA filters to remove airborne contaminants before air is exhausted outside.

3 Is Biological Safety Cabinet Class II Type B2 appropriate for handling TB cultures?

It is recommended for handling Mycobacterium tuberculosis due to its 100% exhaust and strong containment features.

4 Can Biological Safety Cabinet Class II Type B2 be used in animal research labs?

It is useful for animal biosafety when working with infectious agents or volatile substances.

5 Is Biological Safety Cabinet Class II Type B2 recommended for work with aerosol-generating procedures?

It is specifically designed to safely contain aerosols generated during high-risk lab procedures.

6 Is it safe to handle bloodborne pathogens like HIV or Hepatitis B using Biological Safety Cabinet Class II Type B2?

It provides proper containment for bloodborne pathogens, reducing exposure risk during sample handling.

7 Is Biological Safety Cabinet Class II Type B2 suitable for mycology lab?

It is often used in fungal culture where spore control and operator safety are priorities.

8 Is Biological Safety Cabinet Class II Type B2 suitable for emerging infectious disease research?

It is often used in high-containment labs researching new pathogens due to its strict exhaust and filtration protocols.

9 Is Biological Safety Cabinet Class II Type B2 suitable for handling genetically engineered bacteria or viruses?

It offers safe containment for recombinant organisms.

10 Is Biological Safety Cabinet Class II Type B2 suitable for environmental sample testing?

It is suitable for environmental sample testing when pathogens or volatile contaminants are present.