Features:
- Innovated rich oxygen air-acetylene flame analysis technique(WFX-110B):
The patented flame analysis technique adopting rich oxygen air-acetylene flame as the substitution for nitrous oxide-acetylene flame for high temperature element analyses, such as Ca, Al, Ba, W, Mo, Ti, V, etc. Flame temeperature is continuously adjustable between 2300-2950℃ which makes it possible to choose the best atomization temperature for different elements. It features easy operation, low analysis cost and wide flame AAS analytical range. Rich oxygen flame will not pollute the environment and is not harmful to human bodies. It's break-through in flame AAS analysis
- Flame atomization system with flame emission burner:
A flame emission burner head can be installed to perform flame emission analysis to Alkali metals as K, Na etc.
- Accurate fully automated control system:
Automatic multi-lamp turret, automatic adjustment of lamp current and optimization of light beam position
Automatic wavelength scanning and peak picking
Automatic spectral bandwidth changing
Automatic ignition
- Perfect safety protection measures:
Alarm and automatic protection to fuel gas leakage, abnormal flow, insufficient air pressure and abnormal flame extinction in flame system
- Advance and reliable electronic design:
Adopting large-scale programmable logic array and Inter I2C bus technology
European type sockets and AMP adapters with high reliability to ensure long term reliability of the whole electronic system
- Easy and practical analysis software:
Easy-to-use AAS analysis software is made under Windows operating system, realizing fast parameter setting and optimization
Automatic sample dilution, automatic calculation and analytical result automatic print out
Technical Specification:
|
Model No: AAS-011 |
||
|
Main Specification |
|
190-900nm |
|
Wavelength Accuracy |
±0.25nm |
|
|
Resolution |
Two spectral lines of Mn at 279.5nm and 279.8nm can be energy separated with the spectral bandwidth of 0.2nm and valley- peak ratio less than 30% |
|
|
Baseline Stability |
≤0.004A/30min |
|
|
D2 lamp Background Correction |
1A is better than 30 times |
|
|
S-H Background Correction |
1.8A better than 30 times |
|
|
Light Source System |
Lamp Turret |
6-lamp turret, 4-lamp turret ; Auto-alignment, fully automated scan and peak-picking. |
|
Lamp Current Adjustment |
400Hz square wave pulse. Wide pulse current: 0~25mA, Narrow pulse current: 0~10mA |
|
|
Lamp Power Supply mode |
400Hz square wave pulse; 100Hz narrow square wave pulse+400Hz wide square wave pulse |
|
|
Optical System |
Monochromator |
Single beam, Czerny turner design grating monochromator |
|
Grating |
1800 l/mm |
|
|
Focal Length |
277mm |
|
|
Blazed Wavelength |
250nm |
|
|
Spectral Bandwidth |
0.1nm, 0.2nm, 0.4nm, 1.2nm automatic change |
|
|
Flame Atomizer |
Burner |
10cm single slot all-titanium burner |
|
Spray chamber |
Corrosion resistant all-plastic spray chamber |
|
|
Nebulizer |
High efficiency glass nebulizer with metal sleeve, sucking up rate:6-7ml/min |
|
|
Detection and Data Processing System |
Detector |
Photomultiplier with high sensitivity and wide spectral range |
|
Software |
Windows Operating System |
|
|
Analytical Method |
Working curve auto-fitting; standard addition method; automatic sensitivity correction, automatic calculation of concentration and content. |
|
|
Repeat Times |
Maximum 20 times of repeat measurement, automatic calculation of mean value, standdard deviation and relative standard deviation |
|
|
Multi-task Functions |
Sequential measurement for multi-element determination in one sample |
|
|
Condition |
With model function |
|
|
Result Printing |
Measurement of Data and Final Analysis report printout, editing with Excel |
|
|
Standard RS-232 serial port communication |
||
|
Characteristics Concentration and Detection Limit |
Normal Air C2H2 Flame |
Cu: Characteristic concentration≤0.025mg/L, Detection limit |
|
Oxygen Rich Air C2H2Flame |
Ba: Characteristic concentration≤ 0.22mg/L |
|
|
Al: Characteristic concentration ≤ 0.4mg/L |
||
|
Hydride Generation |
Hg: detection limit≤0.5µg/L |
|
|
Function Expansion |
Hydride vapor generator can be connected for hydride analysis |
|
|
Dimension |
Main Unit |
1020 (L) × 492 (W) × 540 (H)mm |
|
Weight |
Main Unit |
80kg , Unpacked |