Atomic Absorption Spectrophotometer AAS-007

    Features:
    • The flame analysis technique adopting oxygen rich air acetylene flame as the substitution for nitrous oxide-acetylene flame for high temperature element analyses.
    • Automatic hollow cathode lamp (HCL) changeover. The high performance HCL can be used directly.
    • Auto-changeover and position optimization of integrated flame/graphite furnace atomizer.
    • Light-controlled graphite furnace power supplier, adopting FUZZY-PID temperature control technique and dual curve working mode, ensures fast heating, accurate and stable temperature control and good temperature reproducibility.  Temperature auto-correction function provided.
    • Graphite furnace with pneumatic control and pressure lock ensures constant pressure and reliable contact.
    • High accuracy and reproducibility of the graphite furnace auto-sampler, realizing fully automation of graphite furnace analysis.
    • Auto-ignition, automatic spectral bandwidth selection.
    • Fully automated wavelength scanning and peaking
    • Automatic setting of Gas flow
    • Perfect safety protection measures, alarm and automatic safety protection to fuel gas leakage, air pressure deficiency and abnormal flame extinction; spray chamber made of explosion proof material, all titanium burner and explosion proof stopper for flame analysis; alarm and protection function to argon pressure deficiency, insufficient cooling water supply, over-heating and over-current etc. in graphite furnace analysis.
    • Adopting large scale programmable logic array and Inter IC technique and highly reliable adapters for electric design.
    • Standard RS-232 serial interface communication
    • Easy and flexible WINDOWS Operating System, with multi-windows processing technology to observe the signal profile, the measured results and the calibration curve at the same time.  Various analytical programs, data, signals and graphics can bestored for later use.
 
 

Technical Specification:

Model No:  AAS-007


Main Specification

Wavelength Range

190-900nm

Wavelength Accuracy

Better than ±0.25nm

Resolution

Two spectral lines of Mn at 279.5nm and 279.8nm can be energy separated with the spectral bandwidth of 0.2nm and valley- peak ratio less than 30%

Baseline Stability

0.004A/30min

D2 lamp Background Correction

1A is better than 30 times

S-H Background Correction

1.8A better than 30 times

Light Source System

Lamp Turret

Motorized 6-lamp turret (Two high performance HCLs can be mounted on the turret to increase the sensitivity in flame analysis.);

Lamp Current Adjustment

Wide pulse current: 0~25mA, Narrow pulse current: 0~10mA  

Lamp Power Supply mode

400Hz square wave pulse;

100Hz narrow square wave pulse+400Hz wide square wave pulse 

Optical System

Monochromator

Single beam, Czerny turner design grating monochromator

Grating

1800 l/mm

Focal Length

277mm

Blazed Wavelength

250nm

Spectral Bandwidth

0.1nm, 0.2nm, 0.4nm, 1.2nm auto-switch over

Flame Atomizer

Burner

10cm single slot all-titanium burner

Spray chamber

Corrosion resistant all-plastic spray chamber

Nebulizer

High efficiency glass nebulizer with metal sleeve, sucking up rate:6-7ml/min

Emission burner provided

 

Graphite Furnace

Temperature range

Room temperature~3000ºC

Heating rate

2000 ºC /s

Graphite tube dimensions

28mm(L)×8mm(OD)

Precision

Cd≤3%, Cu≤3%, Mo≤4%

Detection and Data Processing System

Detector

R928 Photomultiplier with high sensitivity and wide spectral range.

Software

Under Windows Operating System

Analytical Method

Working curve auto-fitting; standard addition method; automatic sensitivity correction, automatic calculation of concentration and content.

Repeat Times

1-99 times, automatic calculation of mean value, standard deviation and relative standard deviation.

Multi-task Functions

Sequential measurement for multi-element determination to the same sample

Condition Reading

With model function

Result Printing

Measurement of  Data and Final Analysis report printout, editing with Excel

Standard RS-232 serial port communication

Graphite Furnace Auto sample

Sample tray capacity

55 sample vessels and 5 reagent vessels

Vessel material

Polypropylene

Vessel volume

3ml for sample vessel, 20ml for reagent vessel

Minimum sampling volume

1μl

Repeatable sampling times

1 -99 times

Sampling system

Accurate dual pump system, with 100μl and 1ml injiectors

Characteristics Concentration and Detection Limit

Normal Air C2H2 Flame

Cu: Characteristic concentration≤0.025mg/L, Detection limit
≤0.006mg/L

Oxygen Rich Air C2H2Flame

Ba: Characteristic concentration≤ 0.22mg/L

Al: Characteristic concentration ≤ 0.4mg/L

Hydride Generation

Hg: detection limit≤0.5µg/L

Function Expansion

Hydride vapor generator can be connected for hydride analysis

Dimension
(L x W x H) mm and Weight

Main Unit

1070 (L) × 490 (W) × 580 (H)mm , 140kg

Graphite furnace

420×420×460mm, 65kg

Auto sampler

400×290×290mm, 15kg

 
 
Accessories :
Air compressor
Hydride vapor generator
Pyrolytically coated platform graphite tube
Pyrolytically coated graphite tube
Normal graphite tube
High effeciency glass nebulizer