Model No: AAS-004
|


|
|
Main Specification
|
Wavelength Range
|
190-900nm
|
|
Wavelength Accuracy
|
±0.5nm
|
|
Resolution
|
better than 0.3nm
|
|
Baseline Stability
|
0.005A/30min
|
|
Background Correction
|
The D2 lamp background correction capability at 1A is better than 30 times. The S-H background correction capability at 1.8A is better than 30 times
|
|
Hollow Cathode Lamps
|
Lamp Turret
|
6-lamp turret automatic changeover
|
|
Lamp Current Adjustment
|
Automatic adjustment and display.Wide pulse current: 0~25mA, Narrow pulse current: 0~10mA.
|
|
Lamp Power Supply mode
|
Powered by two pulses: 400Hz wide pulse and 100Hz narrow pulse simultaneously
|
|
Optical System
|
Monochromator
|
Single beam, Czerny-Turner design grating monochromator
|
|
Grating
|
1800 l/mm
|
|
Focal Length
|
277mm
|
|
Blazed Wavelength
|
250nm
|
|
Spectral Bandwidth
|
0.1nm, 0.2nm, 0.4nm, 1.2nm
|
|
Detection and Data Processing System
|
Detector
|
Photomultiplier with high sensitivity and wide spectral range
|
|
Software
|
Ms-WindowsTM operating system
|
|
Analytical Method
|
Working Curve Method (1st order and 2nd order fitting), Standard addition method
|
|
Repeatable Measurement
|
Maximum 20 times of repeat measurement, automatic calculation of mean value, standard deviation and relative standard deviation
|
|
Condition Reading
|
With model function
|
|
Result Printing
|
Measurement of data and final analytical report printout
|
|
Flame Analysis
|
Burner
|
10cm single slot all titanium burner for Air-C2H2flame; 5cm single slot all titanium burner for oxygen rich flame
|
|
Spray Chamber
|
All plastic spray chamber
|
|
Nebulizer
|
High efficiency glass nebulizer with metal sleeve
|
|
Position Adjustment
|
Adjusting mechanism for vertical, horizontal positions and the rotation angle of the burner
|
|
Protection of Gas System
|
Fuel gas leak protection, abnormal flame extinction alarm, air-pressure deficiency protection and explosion proof device
|
|
Flame Ignition
|
Automatic
|
|
Characteristics Concentration and Detection Limit
|
Normal Air C2H2 Flame
|
Cu: Characteristic concentration ≤0.05mg/L, Detection limit≤0.006mg/L;
|
|
Oxygen Rich Air C2H2Flame
|
Ba: Characteristic concentration ≤0.22mg/L
|
|
Hydride Generation
|
Hg: detection limit ≤ 0.5μg/L
|
|
Dimension (L x W x H) mm
|
1020x490x540
|
|
Weight
|
Unpacked 80kg
|