Features:
- High pressure solvent delivery pump adopting electronic pulsation suppression technology displacing traditional mechanical buffer to thoroughly solve the problem caused by frequent leakage of mechanical buffer, and to reduce the dead volume of solvent delivery system to the minimum.
- The wavelength positioning of Multi-wavelength UV-Vis detector is realized via the grating's turning which is directly controled by the SCM( Single Chip Micyoco). It has taken place the conventional sine-pole mechanics (monochromator), eliminating the error of wavelength positioning brought by the mechanical abrasion and returning error of the Guide-screw transmission, and also avoiding the monochromator's hitting accident which might be brought by false operation.
PC control and Chromatograph data processing system:
- Complete PC control function, realizing parameter control and real-time monitoring to all parts, including column oven.
- Interface language can be switched between Chinese and English, with simple operation.
- Analytical task can be completed in single sampling, with automatic processing of chromatograph data and obtaining perfect spectra.
- Backstage operation function enables data processing to the existing spectra, while analysis is going on.
- The gradient program of the pump and wavelength time-program of the detector can be established.
- The test report format can be flexibly modified.
- Timer turn-off function is provided.
Technical Specification:
|
Model No: HP-001 |
||
|
High Pressure Pump |
Advanced double-piston pump and electronic residual technique, greatly improves the flow rate precision and repeatability of the pump. |
|
|
|
0.001 ~ 9.99 mL/min in 0.001 mL/min increments. |
|
|
Structure |
Double – Pistin pump. |
|
|
Flow Rate Precision |
0.5% |
|
|
Repeatability |
0.1% |
|
|
Pulsatiion |
0.04 MPa |
|
|
Gradient Mode |
a low pressure gradient system up to 4 solvents and a high pressure gradient system up to 2 solvents |
|
|
Remote Control |
RS-232 connector, PC control |
|
|
System Protection |
Soft start, P min and P max adjustment |
|
|
Display |
LCD, 2×16 |
|
|
Dimensions (W×H×D) |
350×190×210mm |
|
|
UV-Vis detector |
Time-programmed wavelength charge function, realizing online wavelength switch. |
|
|
Baseline drift |
15×10-5AU/h at 240nm, t=1.0s |
|
|
Baseline noise |
2×10-5AU at 240nm, t=1.0s |
|
|
Wavelength range |
190~740nm, △ λ≤8nm |
|
|
Wavelength accuracy |
±1nm |
|
|
Time constants |
0.1/0.2/0.5/1.0/2.0/5.0/10.0 |
|
|
Analog output |
±1V |
|
|
Auto-zero range |
full scale |
|
|
Light source |
deuterium lamp |
|
|
Display |
LCD, 2×16 |
|
|
Dimensions (W×H×D) |
350×190×210mm |
|
|
Column Oven |
Highest temperature establishing, overrun temperature alarming. |
|
|
Control mode |
PID |
|
|
Temperature control range |
ambient~100 ºC |
|
|
Temperature accuracy |
±0.3 ºC |
|
|
Safety setting-up |
overrun temperature limit alarming |
|
|
Display |
4 bits LED ( red-actual temperature, green-establishing temperature) |
|
|
Column |
2 |
|
|
Remote control |
RS-232 connector, PC control |
|
|
Inside dimensions (W×H×D) |
50×380×50mm |
|
|
Outside dimensions (W×H×D) |
120×500×230mm
|
|